1700 Series Analyzers
  Transferring Data from 1700 Series'/Intelli-Trend  to AvTrend
 

 

Tip Archive

23x10.gif (829 bytes)

AvTrend is the new ACES Systems' software for displaying, printing, and archiving spectral data from ACES Systems' analyzers.

It was designed specifically for the Model 2000 Series ProBalancers and the  VIPER 4040 analyzer but can also be a replacement for the older DOS version of ACES Systems' software called Intelli-Trend.

AvTrend will import and display the data currently stored in Intelli-Trend databases as well as accept downloads directly from the 1700 Series analyzers.

This Tip of the Month is intended to give you a short "How To" on the dumping of spectral data from the 1700 series analyzers to AvTrend. Follow these simple steps to complete the data dump.

  1. Use ONLY a null model cable or lap link cable for connecting the analyzer to the com (serial communications) port of your computer. You may require a gender changer depending on your analyzer model and computer configuration but no other adapters should be used.
  2. Start AvTrend and left mouse click on OPTIONS | COMMUNICATIONS | 1700.
  3. When the 1700 Serial Port Setting dialog box appears, select the PORT and SPEED settings as required, then click on OK. Be sure the com port you select here is the one you have connected the null modem or Laplink cable to on your computer.
  4. Turn the 1700 series analyzer ON. When the OPERATION OPTIONS menu is displayed, press the SPCL MODES key. From the SPECIAL MODES MENU select "4—Communicate" and press ENTER. From the COMMUNICATIONS menu, select "1—Configure Serial Interface" and press ENTER. When the communications parameters are displayed, match the "Baud " field with the "Speed" field you defined earlier in step three on your computer. The remaining fields should be: none, 8 data, 1 stop, 20, 1 and 1 in that order from top to bottom following the Baud setting.
  5. Press ENTER. When the COMMUNICATIONS screen is displayed again, select "5—Store Data into Remote" and press ENTER.
  6. Immediately go back to the computer and left mouse click on FILE | TRANSFER | RETRIEVE 17XX DATA.
  7. The communications sequence will begin and continue until all data is transferred to the PC. A message will be displayed indicating when the transfer is complete.
  8. If you encounter an error message indicating no communications between the analyzer and the PC, recheck and verify the type cable you are using, the com port you are connected to and the com port selected in AvTrend. You should also verify the Baud rate (speed settings) in both the analyzer and the PC.
  9. When all else fails and you are still unable to communicate, call ACES Systems Product Support or contact us via email at support@acessystems.com.
 

 